Development and Validation of an Eco-Friendly RP-HPLC Method for
the Quantitative Estimation of Desidustat in Human Plasma
Gauri Vispute¹*, Dr. Upasana Yadav²
1.
Department of Pharmaceutical
Analysis, Parul Institute of Pharmacy, Parul University, Vadodara, Gujarat,
India.
2.
Department of Pharmaceutical
Chemistry, Parul Institute of Pharmacy, Parul University, Vadodara, Gujarat,
India.
Correspondence: gaurivispute18@gmail.com
DOI: https://doi.org/10.71431/IJRPAS.2026.5802
|
Article
Information
|
|
Abstract
|
|
Research Article
Received: 08/08/2026
Accepted: 01/09/2026
Published:04/09/2026
Keywords
Desidustat,
Bioanalytical method, HPLC, Human plasma.
|
|
Desidustat is an inhibitor of
hypoxia-inducible factor-prolyl hydroxylase for the treatment of anemia
associated with chronic kidney disease. For the estimation of Desidustat in
human plasma, a precise, accurate, and sensitive RP-HPLC method has been
developed and validated. The separation was achieved using the isocratic
method with an Agilent Zorbax Eclipse C18 column. The separation was
performed using a flow rate of 1mL/min and a detection wavelength of 238 nm.
The retention time of Desidustat and IS were 3.9 and 6.4 minutes,
respectively. The linearity is observed within the concentration range 0.6 –
36 µg/mL, exhibiting a robust correlation coefficient of 0.9999. The validation and stability study were
conducted in accordance with the ICH M10 guidelines. The findings from the
study suggest that the method may apply to pharmacokinetic, bioequivalence,
and therapeutic drug monitoring studies in routine practice.
|
INTRODUCTION
Bioanalysis is a sub-discipline of
analytical chemistry that is concerned with the quantitative analysis of
xenobiotics and endogenous materials in biological matrices (1). Bioanalytical data are invaluable throughout the
whole drug development process, starting with the optimization of the lead
compound and continuing into late-stage clinical trials and post-marketing
surveillance of the drug (1). These studies offer a basic knowledge of the
pharmacokinetics of a drug through the characterization of the absorption,
distribution, metabolism, and excretion of the drug (1). Moreover, bioanalysis plays a vital role in
measuring toxicokinetics, bioavailability, and bioequivalence (BA/BE) and
therapeutic drug monitoring to prevent toxicity in drugs with narrow
therapeutic indices (2,3). All bioanalytical procedures should meet high
international standards of regulatory standards to guarantee the scientific
integrity of clinical data, including the ICH M10 guidelines (4).
Traditionally, renal anemia was treated
with the help of iron supplementation and injectable Erythropoiesis-Stimulating
Agents(9). Nevertheless, ESAs have been linked to the risks
of hypertension and cardiovascular complications when used in
supraphysiological doses (10) . Desidustat is a novel, second-generation,
small-molecule Hypoxia-Inducible Factor Prolyl Hydroxylase inhibitor (11). Desidustat, commercially marketed as Oxemia™ and
also known by the developmental codes ZYAN1 and BCP29692, is a novel
therapeutic agent developed for the management of anemia. Chemically,
Desidustat possesses the molecular formula C₁₆H₁₆N₂O₆ with an average molecular
weight of approximately 332.31 g/mol. The melting point of the reported
crystalline polymorphs of Desidustat lies within the range of 198–200 °C.
Desidustat demonstrates a LogP value of approximately 0.57, suggesting moderate
hydrophilic behavior with balanced lipophilicity. It acts by stabilizing the
subunit of HIF-alpha, as it would be naturally adapted to hypoxia (11–13). This stabilization triggers the expression of
genes involved in the production of EPO over physiological levels and enhances
iron control through a decrease in the expression of hepcidin(14). The DREAM-D and DREAM-ND studies have
demonstrated clinical efficacy, which is non-inferior to traditional injectable
ESAs. It is important to note that Desidustat has the clinical benefit of being
administered orally thrice in a day, which greatly increases patient compliance
(11,15).
Figure 1 Structure of Desidustat
When evaluating the greenness of an
analytical approach, many factors are taken into account, including the
quantity and toxicity of reagents, generated waste, energy requirements, the
number of procedural steps, miniaturization, and automation. An analytical method's
environmental friendliness and safety are evaluated using the AGREE (Analytical
GREEnness) approach(16). The Green Analytical Procedure Index (GAPI),
assesses the environmental impact of the analytical process at various phases
using five different colored pentagons. To build upon GAPI, a new instrument
called the Complementary Green Analytical Method Index (ComplexGAPI) was
created, which adds more fields related to the procedures carried out before
the actual analytical method(17).
Although Desidustat is a therapeutically
significant compound, there is a serious gap in the scientific literature since
no bioanalytical procedures were documented to determine its concentration in
human plasma. Although Liquid Chromatography-Tandem Mass Spectrometry
(LC-MS/MS) is a high-sensitivity technique, its regular clinical application is
constrained by its high cost of operation and special personnel requirement.
Reversed-Phase High-Performance Liquid Chromatography with UV detection, on the
other hand, offers a relatively inexpensive, dependable, and readily available
alternative to routine therapeutic drug monitoring and pharmacokinetic
investigations.
MATERIALS AND METHOD
Chemicals and Reagents
Desidustat and
Enarodustat reference standards were received as gift samples from Swapnroop
Drugs and Pharmaceuticals, India. Blank human plasma was obtained from the Red
Cross, Jalgaon. HPLC-grade methanol, acetonitrile, and orthophosphoric acid
were procured from Merck Ltd., India. During the experiment, water was sourced
from the Milli-Q water purifying system.
Instrumentation
An Agilent 1100 series HPLC system with a
quaternary pump, autosampler, and Diode Array Detector, and controlled by
ChemStation version 10.1 software, was used for analysis. Separation was done by an Agilent Zorbax Eclipse
C18 column (250 mm x 4.6 mm, 5 um particle size).
Stock and Working Solutions Preparation
Standard Stock Solution: 25 mg dissolved in
50mL methanol. A major stock solution of
Desidustat was made at 500 µg/mL in methanol. On the same note, an internal
standard stock solution of Enarodustat was made at a concentration of 500 µg/mL
in methanol.
Calibration Standards: Standard
stock solution was diluted with methanol to make 6, 60, 110, 160, 210, and 360
µg/mL working standard solutions. These were further spiked into blank human
plasma to produce the ultimate calibration range of 0.6, 6, 11, 16, 21, and 36
µg/mL.
Quality
Control Samples: Appropriate volumes (0.12, 1.2, 5.2, and 6.2 mL)
were pipetted into 10 mL volumetric flasks, and the volume was made up to 10 mL
with mobile phase. These working solutions were spiked into blank human plasma
to prepare QC samples with final concentrations. There
were four levels of quality control in human plasma: 0.3 µg/mL, 3 µg/mL, 13
µg/mL and 15.5 µg/mL.
Internal Standard Working Solution: Stock solution of the internal standard was added to methanol to bring
the initial working IS concentration to 5 µg/mL.
Plasma Sample Extraction Process
Sample cleanup was done through a protein
precipitation technique [3,4]. 0.1 mL of human plasma was transferred into a
clean centrifuge tube. 0.1 mL of Desidustat (API) standard solution and 0.1 mL
internal standard (IS) were added to the plasma. The mixture was vortex-mixed
for 1–2 minutes to ensure uniform distribution of the analyte. Protein
precipitation was carried out by adding an appropriate volume of organic
solvent (acetonitrile, 1.7 ml). The samples were vortexed and then centrifuged
at 5000 rpm for 10 minutes to separate precipitated plasma proteins. The
resulting clear supernatant was carefully collected and filtered through a 0.22
µm syringe filter.
The filtered extract was transferred into HPLC
vials and injected into the HPLC system for analysis.
Chromatographic conditions
The chromatographic separation was
carried out by injecting (20 µL) sample into the HPLC system connected to the Agilent Zorbax Eclipse C18 column (250 mm x 4.6mm,
5 µm particle size). It was run in isocratic mode at 1.0 mL/min and ambient column
temperature. The measurement was performed at 238 nm. The mobile phase was
methanol and 0.1% Orthophosphoric acid (pH 3.0).
Bioanalytical Method
Validation
Method
validation was performed to ensure that the developed HPLC method for
quantification of Desidustat in human plasma is reliable, reproducible, and
suitable for routine bioanalytical application. The method validation was carried out according to ICH M10 (2–4).
System suitability
System suitability was evaluated before
analysis using six replicate injections of the system suitability sample.
Retention time, theoretical plate number, tailing factor, resolution, and
peak-area precision were assessed. Six replicate injections of the 36 µg/mL system suitability sample were
analyzed (n = 6). The ICH parameters for system
suitability were calculated. Retention time (Rt), theoretical plates, and
tailing factor were considered for system suitability.
Linearity
Linearity
of the developed HPLC method was evaluated to establish the relationship
between concentration and detector response for Desidustat in human plasma. The calibration curve was plotted using six data
points by introducing accurate volumes of a working solution into a blank
plasma sample. The ultimate concentration of 0.6 to 36 µg/mL was prepared, and the peak area was noted to plot
the calibration curve. The calibration curve was plotted as peak area versus
concentration. Evaluated on a six-point calibration curve (0.6 to 36 µg/mL).
Precision
Bioanalytical accuracy and precision follow strict FDA and ICH M10
guidelines. Accuracy and precision must be evaluated at a minimum of 4
concentration levels. It is determined
by examining five replicates of the four levels of QC on three consecutive days
to determine intra-day and inter-day performance.
Accuracy and precision
The accuracy and precision of the developed bioanalytical HPLC method
for Desidustat were evaluated to assess the closeness of measured values to
true values and the method's repeatability. For the recovery study, the peak
areas of extracted plasma samples of Desidustat at LQC, MQC, and HQC levels
were determined. Recovery samples at LQC, MQC, and HQC were prepared using
pre-prepared plasma quality control stocks and API standard solutions, without
employing the standard addition method. Intra-day accuracy and precision were
evaluated by analyzing QC samples at different concentration levels within the
same day. Each QC level was analyzed in three replicates. Inter-day accuracy
and precision were evaluated by analyzing QC samples on different days.
The determination of analyte recovery was
conducted by comparing the analytical outcomes of extracted samples. The area
of the extracted plasma samples with unextracted aqueous standards at LQC, MQC,
and HQC levels.
Robustness
To
evaluate the robustness of the developed HPLC method for Desidustat by introducing
small, deliberate variations in chromatographic conditions and observing their
effect on method performance. The analysis was performed at a fixed
concentration of 3 µg/mL.
Stability
The assessment of stability was carried out as
an essential At LQC and HQC levels, stability was investigated under different
conditions, such as bench-top and long-term stability, by comparing the results
with freshly prepared ones
Carryover and dilution integrity
Carryover was determined by injecting blank
plasma following the Upper Limit of Quantification sample; dilution integrity
was determined by injecting a 36 0g/mL sample with blank plasma; and robustness
was measured by making small, intentional changes in flow rate (10%),
wavelength (2nm), and mobile phase composition.
Green characteristics of the method
Numerous
approaches such as AGREE, GAPI, complex GAPI, and others have been developed to
evaluate the environmental sustainability of analytical procedures. AGREE and
GAPI use a color-coded system to assess the environmental friendliness of the
method. Red indicates that the method is not eco-friendly, yellow suggests that
the method may be somewhat eco-friendly, and green signifies that the method is
environmentally friendly.
The
AGREE score is shown in the center of the pictogram; scores below 0.50 indicate
that the approach is not acceptable or environmentally friendly; scores between
0.50 and 0.70 indicate the method is suitable, and scores greater than 0.75
indicate the method developed is highly effective in terms of environmental
sustainability. The AGREE pictogram displays three colors with varying
intensities depending upon the inputs provided. The pale green color in AGREE
indicates sustainability, safety, and reduced waste generation. Green indicates
that the method is eco-friendly (16,18).
The expanded version of GAPI, Complex GAPI was
introduced in 2021. Each hexagon in a GAPI figure, whose color intensity varies
with the data entered, represents an individual aspect of the process. GAPI’s primary goal is to quantify green
criteria into a pictogram by assessing variables including waste, energy,
chemicals, and risks. From sample collection to final analysis, GAPI evaluates
an analytical method's overall eco-friendliness. The primary benefit of Complex
GAPI get a chance to access the pre-analytical process parameters, including
purification techniques, reagents and solvents, instrumentation and sensor
preparation (19).
RESULT AND DISCUSSION
Development of a
chromatographic method
Solubility
studies indicated high dissolution in acetonitrile and methanol, whereas only
moderate solubility was observed in water and isopropanol. Based on these
findings, methanol was selected as the solvent for preparing both stock and
diluted solutions. The observed melting range of 198–202 °C closely aligned
with reported values, indicating a high degree of sample purity. This
consistency further supported the reliability of the material throughout the
experiments. During UV–visible analysis, a distinct peak absorbance was
observed at 238 nm. Accordingly, this wavelength was selected for HPLC
detection, as it provided a strong signal response without interference from
methanol.
Method
development for the RP-HPLC analysis began with a systematic evaluation of
different mobile phase compositions to achieve effective separation of
Desidustat. Each trial mixture was carefully assessed, with particular emphasis
on peak symmetry as a key indicator of performance. The inclusion of
Enarodustat as an internal standard enabled consistent monitoring throughout
the optimization process. Gradual, stepwise refinement ultimately led to the
establishment of optimal chromatographic conditions.
Initial
trials using a strong mobile phase composition of methanol and 0.1%
orthophosphoric acid (70:30, v/v) resulted in rapid elution but poor peak
resolution. Major Desidustat peak at ~4.20
min (area ~130.7 mAU·s), small late peak at ~5.46 min. Baseline remains stable;
peaks are reasonably sharp. Mobile-phase change
shifted RT slightly (~4.2 min) and produced an acceptable peak shape. Rejected
the mobile phase because of poor resolution.
Adjusting the composition to a more balanced
ratio, with reduced organic content, led to slight improvements in separation;
however, issues such as peak asymmetry and baseline disturbances persisted. A
mobile phase consisting of methanol and 0.1% orthophosphoric acid (pH 3.0) in a
30:70 (v/v) ratio provided superior results, yielding well-resolved and
symmetrical peaks. The 30:70 (v/v) methanol:0.1% orthophosphoric acid
composition provided the most favorable chromatographic performance among the
evaluated conditions, based on retention, peak symmetry, and resolution. Under
optimized parameters, a clear separation of Desidustat and the internal
standard was achieved within 8 minutes. This performance was facilitated by a
Zorbax Eclipse C18 column, whose stationary-phase properties enabled effective
interaction with and resolution of these moderately water-soluble compounds in
plasma samples.
The
mobile phase was optimized to a composition of methanol and 0.1%
orthophosphoric acid (pH 3.0) in a 30:70 (v/v) ratio and was operated in
isocratic mode at a flow rate of 1.0 mL/min. The injection volume was set at 20
µL. Detection was carried out at 238 nm, corresponding to the λmax of
Desidustat as determined from preliminary UV–visible spectrophotometric
analysis. The total run time was 8 minutes, with Desidustat and Enarodustat
eluting at approximately 3.9–4.0
minutes and 6.3–6.4 minutes, respectively. Optimized chromatographic
conditions are shown in Table 1.
Table 1 Optimized
Chromatographic Conditions for HPLC Method Development
|
Parameter
|
Condition
|
|
Column
|
Zorbax Eclipse C18 column (250 × 4.6 mm, 5 µm)
|
|
Particle Size
|
5 µm
|
|
Mobile Phase
|
Methanol: 0.1% orthophosphoric acid (30:70, v/v)
|
|
Mode
|
Isocratic
elution
|
|
Flow Rate
|
1.0 mL/min
|
|
Detection Wavelength
|
238 nm
|
|
Injection Volume
|
20 µL
|
|
Column Temperature
|
Ambient
|
|
Run Time
|
8 minutes
|
Figure 2:
UV spectra of Desidustat
Figure 3:
Chromatogram of Desidustat and IS from plasma matrix
Bioanalytical Method
Validation
A
key step in the development of drugs and biological products is the
quantitative assessment of drug concentration in biological matrices (such as
serum, plasma, blood, urine, and saliva). In the pharmaceutical sector,
bioanalysis is crucial since it aids in the quantitative assessment of both
high molecular weight substances like proteins and low molecular weight
medications. Critical judgments supporting the safety and efficacy of a
medicinal therapeutic ingredient or product depend on the results of several
animal toxicity research, pharmacokinetic studies, and preclinical PK
investigations, including bioavailability and bioequivalency studies. As a
result, validation guarantees that the specified bioanalytical method yields
sensitive and repeatable data.
Validation
studies confirmed that the method complied with ICH M10 guidelines,
demonstrating its reliability for routine clinical analysis. Consistent
performance was observed across varying conditions throughout all evaluation
phases. Given the critical importance of accuracy in testing, each step was
conducted in strict adherence to established protocols. Alignment with
peer-reviewed procedures ensured clarity and reproducibility of results, while
even minor deviations were carefully monitored during repeated trials.
System suitability
System
suitability ensures the instrument and column are performing together to
generate accurate, reproducible data before and during sample analysis. Table 2
shows the system suitability parameters of desidustat and IS in human plasma.
Table 2: System suitability parameters of desidustat and IS in human plasma
|
Sr No
|
Parameter
|
Desidustat
|
Enarodustat (IS)
|
Acceptance Criteria
|
Guideline
|
|
1
|
Retention
Time (Rt)
|
3.9
min
|
6.4
min
|
Consistent
(±2%)
|
ICH
M10
|
|
2
|
Theoretical
Plates (N)
|
4965
|
7557
|
≥
2000
|
ICH
M10
|
|
3
|
Tailing
Factor
|
0.91
|
0.96
|
≤
2.0
|
ICH
M10
|
|
4
|
Resolution
|
8.09
|
9.52
|
≥
2.0
|
ICH
M10
|
|
5
|
%RSD
of Peak Area
|
0.18%
|
1.03%
|
≤
2%
|
ICH
M10
|
Linearity
The method exhibited a strong linear relationship between concentration
and peak area over the range of 0.6 to 36 µg/mL. The regression equation was
determined to be y = 69.51x – 7.897 with an exceptional correlation coefficient
(r2) of 0.9999. The plotted data is of peak area versus
concentration. Table 3 provides linearity data of desidustat, and Figure 4
shows the linearity curve of desidustat
Figure 4:
Overlay of linearity
Figure 5:
Linearity curve of desidustat in human plasma
Table 3: Regression and Linearity Parameters
|
Parameter
|
Result
|
|
Linearity Range
|
0.6 – 36 µg/mL
|
|
Regression Equation
|
y = 69.51x – 7.897
|
|
Correlation Coefficient (R2)
|
0.9999
|
|
Slope
|
69.51
|
|
y-intercept
|
-7.897
|
Accuracy and Precision
The
extent to which an analytical method's mean test findings closely resemble the
analyte's actual value (concentration) is known as its accuracy. When an
analytical method is applied repeatedly to several aliquots of a single
homogeneous volume of biological matrix, the precision of the method describes
the closeness of individual analyte measurements. At least five determinations
per concentration should be used to gauge precision. At least five
determinations per concentration should be used to gauge precision.
It
is advised to use at least three concentrations within the anticipated
concentration range. With the exception of the LLOQ, where it should not exceed
20% of the coefficient of variation (CV), the precision calculated at each
concentration level should not exceed 15% of the CV. Precision is further
separated into two categories: between-run, inter-batch precision or
repeatability, which measures precision over time and may involve various
analysts, tools, reagents, and labs, and within-run, intra-batch precision or
repeatability, which evaluates precision during a single analytical run.
It
is advised to use at least three concentrations within the anticipated
concentration range. With the exception of the LLOQ, where it should not exceed
20% of the coefficient of variation (CV), the precision calculated at each
concentration level should not exceed 15% of the CV. Precision is further
separated into two categories: between-run, inter-batch precision or
repeatability, which measures precision over time and may involve various
analysts, tools, reagents, and labs, and within-run, intra-batch precision or
repeatability, which evaluates precision during a single analytical run.
Replicate
examination of samples with known analyte concentrations determines accuracy.
At
each of the four QC concentrations - LLOQ, LQC, MQC, and HQC - the method
delivered results close to the true values, showing strong reliability.
Performance remained consistent across repeated measurements, indicating low
variability under controlled conditions.
Intra-day
plus inter-day accuracy showed results between 96.61% and 102.86%. These fall
comfortably inside the accepted range of 85–115%. Although slight variations
occurred, none breached minimum standards. Performance stayed consistent across
testing periods. The data met the criteria without exception. Table 4 shows
Intra-day Accuracy and Precision of Desidustat in Human Plasma
Measured
variation stayed tightly within 1.14% across tests, while at the lowest
detectable level it narrowed further, reaching just 0.32%. This reflects
strong consistency when repeated under similar conditions. Intra-day accuracy
and precision were within acceptable limits as per ICH M10 guidelines
Table 4: Intra-day Accuracy and Precision of
Desidustat in Human Plasma
|
QC Level
|
Conc
|
Mean Area
|
Amt Found
|
% Accuracy
|
SD
|
%RSD
|
|
LLOQ
QC
|
0.3
|
13.09
|
0.30
|
100.65
|
0.04
|
0.32
|
|
LQC
|
3
|
206.54
|
3.09
|
102.86
|
2.35
|
1.14
|
|
MQC
|
13
|
902.13
|
13.09
|
100.71
|
0.90
|
0.10
|
|
HQC
|
15.5
|
1082.27
|
15.68
|
101.19
|
0.05
|
0.13
|
Table 5: Inter-day accuracy and precision of desidustat in human
plasma
|
QC Level
|
Conc
|
Mean Area
|
Mean Conc Found
|
% Accuracy
|
SD
|
%RSD
|
|
LLOQ
QC
|
0.3
|
13.14
|
0.30
|
100.90
|
0.04
|
0.32
|
|
LQC
|
3
|
204.10
|
3.05
|
101.66
|
0.28
|
0.14
|
|
MQC
|
13
|
898.45
|
13.04
|
100.30
|
0.38
|
0.04
|
|
HQC
|
15.5
|
1032.94
|
14.97
|
96.61
|
1.17
|
0.11
|
Accuracy
The
recovery was evaluated at low, medium, and high QC levels to assess the
consistency and efficiency of the extraction procedure. The % coefficient of
variation (%CV) for recovery at each QC level was required to be not more than
15%. Accuracy expressed as % Recovery.
|
Sr.no
|
Conc
|
Extracted Area
|
Unextracted Area
|
Amount Found
|
Mean % Recovery
|
%RSD
|
|
L-1
|
3
µg/ml
|
208.76
|
242.185
|
0.86
|
85.17%
|
1.20
|
|
L-2
|
3 µg/ml
|
205.4456
|
242.3868
|
0.84
|
|
L-3
|
3
µg/ml
|
205.4266
|
242.971
|
0.84
|
|
M-1
|
13 µg/ml
|
900.8414
|
931.8355
|
0.96
|
96.83%
|
0.18
|
|
M-2
|
13
µg/ml
|
902.1176
|
930.8342
|
0.96
|
|
M-3
|
13 µg/ml
|
903.4453
|
932.335
|
0.96
|
|
H-1
|
15.5µg/ml
|
1081.5272
|
1120.125
|
0.96
|
95.15%
|
2.66
|
|
H-2
|
15.5µg/ml
|
1081.4542
|
1118.667
|
0.96
|
|
H-3
|
15.5µg/ml
|
1034.8492
|
1122.105
|
0.92
|
Table 6: Accuracy results
To
evaluate the robustness of the developed HPLC method for Desidustat by
introducing small, deliberate variations in chromatographic conditions and
observing their effect on method performance. Even when the flow rate shifted
slightly - by 10% - the approach held steady. Detection wavelength altered by 2
nm, yet results stayed consistent. Mobile phase makeup varied, but performance
did not waver. In every case, %RSD stayed under 2%. Stability emerged clearly
across all tested conditions.
Table 7: Robustness Parameter
|
Parameters
|
Concentration
|
Mean Area
|
SD
|
RSD
|
|
Flow
rate
|
|
|
|
|
|
0.9mL/min
|
3
µg/mL
|
240.20
|
1.54
|
0.64
|
|
1.1mL/min
|
3
µg/mL
|
241.42
|
2.94
|
1.2
|
|
Wavelength
(nm)
|
|
|
|
|
|
237
|
3
µg/mL
|
207.10
|
0.42
|
0.20
|
|
239
|
3
µg/mL
|
217.14
|
0.64
|
0.29
|
|
Mobile
phase ratio
|
|
|
|
|
|
29:71(MeOH:Buffer)
|
3
µg/mL
|
216
|
2.99
|
1.38
|
|
31:69(MeOH:Buffer)
|
3
µg/mL
|
214.72
|
1.12
|
0.52
|
Carryover
Following a
high-concentration injection, no notable carryover appeared in blank plasma,
where the signal reached just below the LLOQ level. Response levels remained
minimal despite prior exposure to concentrated samples. The absence of
lingering effects suggests limited residue transfer during analysis. Signal
detection stayed close to baseline under these conditions.
For
samples exceeding the calibration range (36 µg/mL), the method maintained an
accuracy of 95.16% and a %RSD of 0.15 after dilution with blank plasma. This
confirms the method's validity for analyzing high-concentration clinical
samples.
Green characteristics of the method
AGREE score generates a circular clock-like
pictogram with a central score and color code like green, light green, yellow,
and red with 12 distinct assessments. The optimized HPLC method shows an AGREE
score of 0.8 (Figure 6a), which reflects higher compliance with green
analytical chemistry principles. Organic modifiers like acetonitrile lower the
score due to safety and waste principles.
A
complex GAPI shows a standardized visual pictogram evaluating environmental
sustainability. It evaluates the standard 15 steps of an analytical method from
the original GAPI tool. Extra hexagonal specifically added in the Complex GAPI
framework to incorporate pre-laboratory operations, sample collection,
preservation, and transport(20). In this study, the Complex GAPI score was
found to be 77 (Figure 6b) using parameters such as mobile phase, sample
collection, storage, toxic solvents, and energy consumption.
a: AGREE score b:
Complex GAPI
Figure 6: Greenness values of proposed analytical method
CONCLUSION
An
RP-HPLC method was developed and validated for the quantification of Desidustat
in human plasma. The method uses conventional HPLC-UV instrumentation and
therefore provides a potentially accessible alternative for laboratories
without LC-MS/MS capability. Chromatographic analysis was performed using an
Agilent Zorbax Eclipse C18 column, with a mobile phase consisting of methanol
and 0.1% orthophosphoric acid (30:70, v/v) delivered at a constant flow rate of
1.0 mL/min. Sample preparation was simplified through protein precipitation
using acetonitrile, avoiding complex extraction procedures. Reliable separation
was achieved through careful control of flow rate and mobile phase composition.
The
method demonstrated excellent linearity over the studied range. Accuracy was
within 96.61–102.86%, complying with ICH M10 guidelines, while precision
remained robust with %RSD values ≤1.14%. Recovery was consistent, ranging from
85% to 97% across experiments.
Stability
studies confirmed that the analyte remained unaffected under various storage
conditions, indicating resilience within the plasma matrix. Owing to its
sensitivity and reproducibility, the method is well-suited for monitoring drug
concentrations over time, particularly in patients with renal impairment.
The
method is assessed for its greenness using the AGREE approach and Complex GAPI,
and proved the greenness of the method. A high-performance liquid
chromatography (HPLC) method's greenness assessment has the advantage of
systematically identifying and minimizing environmental pollution, reducing
hazardous chemical waste, lowering energy consumption, and safeguarding
operator safety.
REFERENCES
1. Pandey S, Pandey P, Tiwari G, Tiwari R.
Bioanalysis in drug discovery and development. Pharm Methods. 2010;1(1):14.
doi:10.4103/2229-4708.72223 PubMed PMID: 23781412.
2. Bouchafra H, Orche A El, El Khabbaz C,
Cheikh A, El Karbane M, El M, et al. Determination and validation of
tiaprofenic acid in human plasma: A detailed LC-MS/MS-based analysis following
ICH M10 guidelines and the accuracy profile approach. Current Chemistry
Letters. 2024;13:707–16. doi:10.5267/j.ccl.2024.4.003
3. Bioanalysis of Small-Molecule Drugs and
Metabolites in Physiological Samples by LC–MS, Part 1: An Overview | LCGC
International [Internet]. [cited 2026 Jun 22]. Available from:
https://www.chromatographyonline.com/view/bioanalysis-of-small-molecule-drugs-and-metabolites-in-physiological-samples-by-lc-ms-part-1-an-overview
4. International Council For Harmonisation Of
Technical Requirements For Pharmaceuticals for Human Use ICH Harmonised
Guideline Bioanalytical Method Validation And Study Sample Analysis M10.
5. Chen TK, Knicely DH, Grams ME. Chronic
Kidney Disease Diagnosis and Management: A Review. JAMA. 2019 Oct
1;322(13):1294. doi:10.1001/JAMA.2019.14745 PubMed PMID: 31573641.
6. Ganz T, Nemeth E. Iron Balance and the Role
of Hepcidin in Chronic Kidney Disease. Semin Nephrol. 2016 Mar 1;36(2):87–93.
doi:10.1016/J.SEMNEPHROL.2016.02.001 PubMed PMID: 27236128.
7. Tien C, Meng CC. Anemia of Chronic Kidney
Disease. The Kidney Book: A Practical Guide on Renal Medicine. 2024 Jul
23;137–42. doi:10.1142/9789811283987_0013 PubMed PMID: 30969693.
8. Badura K, Janc J, Wąsik J, Gnitecki S,
Skwira S, Młynarska E, et al. Anemia of Chronic Kidney Disease—A Narrative
Review of Its Pathophysiology, Diagnosis, and Management. Biomedicines 2024,
Vol 12,. 2024 May 27;12(6). doi:10.3390/BIOMEDICINES12061191 PubMed PMID:
38927397.
9. Biggar P, Kim GH. Treatment of renal
anemia: Erythropoiesis stimulating agents and beyond. Kidney Res Clin Pract.
2017 Sep 1;36(3):209. doi:10.23876/J.KRCP.2017.36.3.209 PubMed PMID: 28904872.
10. Rao SRGG, Bugazia S, Dhandapani TPM, Tara A,
Garg I, Patel JN, et al. Efficacy and Cardiovascular Adverse Effects of
Erythropoiesis Stimulating Agents in the Treatment of Cancer-Related Anemia: A
Systematic Review of Randomized Controlled Trials. Cureus. 2021 Sep
9;13(9):e17835. doi:10.7759/CUREUS.17835 PubMed PMID: 34527499.
11. Joharapurkar A, Pandya V, Patel H, Jain M,
Desai R. Desidustat: a novel PHD inhibitor for the treatment of CKD-induced
anemia. Frontiers in Nephrology. 2024;4:1459425.
doi:10.3389/FNEPH.2024.1459425 PubMed PMID: 39502472.
12. Dhillon S. Desidustat: First Approval. Drugs.
2022 Jul 1;82(11):1207–12. doi:10.1007/S40265-022-01744-W PubMed PMID:
35834123.
13. Katekhaye V, Tolani P, Pasari S, Zunke P, Kashiv
P, Dubey S. Desidustat – An oral erythropoiesis stimulator: Pharmacology and
current evidence. Journal of The Nephrology Society. 2022;1(2):79.
doi:10.4103/JTNS.JTNS_24_22
14. Wang Y, Yu X. Stabilizing Hypoxia-Inducible
Factor to Manage Anemia in Chronic Kidney Disease: From Basic Theory to
Clinical Study. Kidney Diseases. 2024 Jan 3;10(2):132. doi:10.1159/000536039
PubMed PMID: 38659701.
15. Gang S, Khetan P, Varade D, Chinta VR, Mavani
S, Gupta U, et al. Desidustat in Anemia due to Dialysis-Dependent Chronic
Kidney Disease: A Phase 3 Study (DREAM-D). Am J Nephrol. 2022 Jun 1;53(5):343.
doi:10.1159/000523949 PubMed PMID: 35462369.